A Wireless Intraocular Pressure Sensor with Variable Inductance Using a Ferrite Material

Citations

WEB OF SCIENCE

18
Citations

SCOPUS

21

초록

A wireless intraocular (IOP) pressure sensor based on micro electro mechanical system (MEMS) technology is proposed. The proposed IOP sensor uses variable inductance according to the external pressure. The proposed sensor is composed of two flexible membranes: a ferrite bottom part, an inductor, and a capacitor. The inductance of the sensor varies according to the external pressure. The resonance frequency of the sensor is also varied, and this frequency is detected using an external coil. The external coil is designed with an FR-4 printed circuit board. The feasibility of the proposed sensor structure using variable inductance to detect the external pressure is successfully demonstrated.

키워드

Ferrite; IOP sensor; resonance frequency; variable inductance
제목
A Wireless Intraocular Pressure Sensor with Variable Inductance Using a Ferrite Material
저자
Kang, Byungjoo; Hwang, Hoyong; Lee, Soo Hyun; Kang, Ji Yoon; Park, Joung-Hu; Seo, Chulhun; Park, Changkun
DOI
10.5573/JSTS.2013.13.4.355
발행일
2013-08
유형
Article
저널명
JOURNAL OF SEMICONDUCTOR TECHNOLOGY AND SCIENCE
권
13
호
4
페이지
355 ~ 360