Development of an All-in-One Phantom and Scintillator Radiation Sensor for Real-Time Monitoring of Source Position and Dose Distribution in High-Dose-Rate Brachytherapy

  • Shin, Sang Hun; 
  • Yoo, Wook Jae; 
  • Jang, Kyoung Won; 
  • Cho, Seunghyun; 
  • Kim, Kum Bae; 
  • 외 1명
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

2

초록

In this study, we developed an all-in-one phantom and scintillator radiation sensor (PSRS) with plastic scintillating fibers which can measure the real-time position and the dose distribution of Ir-192 as a radioactive source employed in high-dose-rate (HDR) brachytherapy. As experimental results, we simultaneously obtained information on the shape and position of the Ir-192 as well as the dose distribution. The all-in-one PSRS system has many dosimetric advantages, such as real-time monitoring, wide dose-response range, high spatial resolution, nearly water-equivalence for high-energy photons, and the ability to measure two-dimensional dose distribution.

키워드

radiation sensor; HDR brachytherapy; dose distribution; position; real-time monitoring; ENERGY BEAM DOSIMETRY; DETECTORS
제목
Development of an All-in-One Phantom and Scintillator Radiation Sensor for Real-Time Monitoring of Source Position and Dose Distribution in High-Dose-Rate Brachytherapy
저자
Shin, Sang Hun; Yoo, Wook Jae; Jang, Kyoung Won; Cho, Seunghyun; Kim, Kum Bae; Lee, Bongsoo
발행일
2016-06
유형
Article; Proceedings Paper
저널명
Sensors and Materials
권
28
호
6
페이지
619 ~ 624