Characterization of the fiber-optic goniometer for measuring knee joint angle

  • Kim, S.G.; 
  • Shin, S.H.; 
  • Jang, K.W.; 
  • Jeon, H.; 
  • Jang, J.S.; 
  • ... Cho, S.; 
  • 외 5명

초록

Gait disturbance is typical symptom of the degenerative cerebropathia such as Alzheimer's disease, dementia, and Parkinson's disease. Thus measuring the activity of the knee joint is important to make a diagnosis for these diseases. In this study, the fiber-optic goniometer was developed using a plastic optical fiber, a light emitting diode, and a photodiode to measure knee joint angles. To evaluate the fiber-optic goniometer, we measured knee angles using the fiber-optic goniometer and 3-D IR motion camera in walking and cycling. As a result, patterns of knee joint angle obtained by the fiber-optic goniometer during walking and cycling corresponded with those obtained by the 3-D IR motion camera. In terms of sampling resolution and error occurrence, the fiber-optic goniometer fabricated through this study can be used in diagnosis of degenerative cerebropathia. Copyright © 2015 American Scientific Publishers All rights reserved.

제목
Characterization of the fiber-optic goniometer for measuring knee joint angle
저자
Kim, S.G.; Shin, S.H.; Jang, K.W.; Jeon, H.; Jang, J.S.; Kim, J.S.; Kwon, G.; Yoo, W.J.; Cho, S.; Song, T.; Lee, B.
DOI
10.1166/sl.2015.3481
발행일
2015
개최국가
미국