습도에 따른 저가형 광산란 기기 초미세먼지 측정 오차 분석
- 김민주;
- 이정훈;
- 우욱용;
- 최하진
초록
Inaccuracy of particulate matter from light scattering sensors has been reported, where the environmental parameters including aerosol size, physiochemical properties, temperature and relative humidity play important roles in the accuracy of measurements. We explored the effect of relative humidity using six light-scattering devices in environmental chamber. Potassium chloride aerosol was exposed to the chamber, and the devices including tree firstclass certified, two out of class, and one dehumidification device, were applied to measure particulate matter. The confidence interval of the data was monitored based on RPj, which is a precision index used in EPA's field sampling procedure and the criteria of the interval is 7%. In both the low-humidity and high-humidity experiments, the measurements from first-class devices were problematic compared to the other devices, where the confidence interval was higher than the 7%. In addition, the results show that the confidence interval did not meet the criteria in highhumidity compared to low–humidity. Therefore, the measurements of particulate matter should be re-considered as their precision, which is not absolute value of the fine dust but is a relative trend.
키워드
- 제목
- 습도에 따른 저가형 광산란 기기 초미세먼지 측정 오차 분석
- 제목 (타언어)
- Evaluation of Low-cost Light Scattering Devices for Ultrafine Dust based on Relative Humidity
- 저자
- 김민주; 이정훈; 우욱용; 최하진
- 발행일
- 2022-06
- 저널명
- 한국생활환경학회지
- 권
- 29
- 호
- 3
- 페이지
- 271 ~ 280