Resonance frequency effects of an external coil for IOP sensor applications

  • Hwang, Hoyong; 
  • Kang, Byungjoo; 
  • Lee, Soo Hyun; 
  • Jang, Cheol-In; 
  • Kang, Ji Yoon; 
  • ... Park, Changkun
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

3

초록

In this study, we investigate the resonant frequency effects of an external coil for a wireless intraocular pressure (IOP) monitoring system. The phase difference of an IOP sensor is improved using the resonant frequency of the external coil. By means of phasor diagrams, we show that the phase difference is improved as the resonant frequency of the external coil approaches the resonant frequency of the IOP sensor. The IOP sensor and a multi-layered external coil are designed using microfabrication technology and an FR-4 printed circuit board, respectively. The resonant frequency of the external coil is varied using an additional capacitor and by changing the number of layers of the multi-layered inductor pattern. From the measured results, we successfully verify that the resonant frequency of the external coil improves the phase difference of the IOP sensor.

키워드

intraocular pressure; MEMS; external coil; phasor diagram; sensor; resonance frequency; INTRAOCULAR-PRESSURE SENSOR; READOUT; SYSTEM
제목
Resonance frequency effects of an external coil for IOP sensor applications
저자
Hwang, Hoyong; Kang, Byungjoo; Lee, Soo Hyun; Jang, Cheol-In; Kang, Ji Yoon; Park, Changkun
DOI
10.1080/09205071.2014.891951
발행일
2014-05-03
유형
Article
저널명
Journal of Electromagnetic Waves and Applications
권
28
호
7
페이지
790 ~ 801