상세 보기
Measurements of planar and depth dose distributions using a scintillating fiber-optic image sensor system for dosimetry in radiotherapeutic applications
- Shin, Sang Hun;
- Yoo, Wook Jae;
- Seo, Jeong Ki;
- Han, Ki-Tek;
- Jeon, Dayeong;
- ... Cho, Seunghyun;
- 외 3명
WEB OF SCIENCE
1SCOPUS
2초록
We developed a scintillating fiber-optic image sensor system (SFISS) using square plastic optical fibers (POFs), a scintillating film, a right-angle mirror, and a complementary metal-oxide semiconductor (CMOS) image camera for dosimetry in radiotherapeutic applications. In this study, the scintillating light images were obtained for measuring two-dimensional planar dose distributions of a 6MV photon beam in a solid-water phantom. We also measured the percentage depth doses (PDDs) of 6 and 15MV photon beams using the SFISS and compared them with those obtained using conventional dosimetry films. The proposed sensor has many advantages, such as real-time readout, high-resolution measurement, and lack of corrections for temperature, pressure, and humidity. From the results of this study, it is expected that a SFISS can be developed to accurately measure the dose distribution in a small beam field for stereotactic radiosurgery (SRS).
키워드
- 제목
- Measurements of planar and depth dose distributions using a scintillating fiber-optic image sensor system for dosimetry in radiotherapeutic applications
- 저자
- Shin, Sang Hun; Yoo, Wook Jae; Seo, Jeong Ki; Han, Ki-Tek; Jeon, Dayeong; Jang, Kyoung Won; Sim, Hyeok In; Cho, Seunghyun; Lee, Bongsoo
- 발행일
- 2013-03
- 유형
- Article
- 저널명
- Optical Review
- 권
- 20
- 호
- 2
- 페이지
- 178 ~ 181