Feasibility Study on Fiber-Optic Goniometer for Measuring Knee Joint Angle

  • Kim, Seon Geun; 
  • Jang, Kyoung Won; 
  • Yoo, Wook Jae; 
  • Shin, Sang Hun; 
  • Cho, Seunghyun; 
  • 외 1명
Citations

WEB OF SCIENCE

9
Citations

SCOPUS

11

초록

In this study, we fabricated a fiber-optic goniometer using a plastic optical fiber, a light-emitting diode, and a photodiode. The cladding of the plastic optical fiber was removed at 0.5-3mm intervals regularly to increase the bending loss of the plastic optical fiber. Also, the output voltages of the photodiode based on light intensity that measured using the fiber-optic goniometer were measured light intensities using the fiber-optic goniometer were measured to evaluate the linearity of the fiber-optic goniometer. Finally, we measured the responses of the fiber-optic goniometer for gait speeds of 3, 5, and 10 km/h using a data acquisition board and a LabVIEW program. (C) 2014 The Japan Society of Applied Physics

키워드

fiber-optic sensor; goniometer; knee joint angle; OPTICAL-FIBER
제목
Feasibility Study on Fiber-Optic Goniometer for Measuring Knee Joint Angle
저자
Kim, Seon Geun; Jang, Kyoung Won; Yoo, Wook Jae; Shin, Sang Hun; Cho, Seunghyun; Lee, Bongsoo
DOI
10.1007/s10043-014-0112-z
발행일
2014-09
유형
Article
저널명
Optical Review
권
21
호
5
페이지
694 ~ 697