상세 보기
Feasibility study on development of Cerenkov fiber-optic dosimeter for radiotherapy application
- Yoo, W.J.;
- Shin, S.H.;
- Han, K.-T.;
- Jeon, D.;
- Hong, S.;
- ... Cho, S.;
- 외 5명
초록
To obtain real-time dose information in photon-beam therapy using a clinical linear accelerator, we fabricated a novel Cerenkov fiber-optic dosimeter using two plastic optical fibers without employing a scintillator. In this study, the light intensity and spectrum of Cerenkov radiation induced by a high-energy photon beam were measured as functions of the irradiation angle and the length difference between the two plastic optical fibers in the dosimeter probe. Also, we obtained a percentage depth dose curve for a 6 MV photon beam with a field size of 10 × 10 cm2 according to the depth of the solid water phantom. Based on the results of this study, it is anticipated that the proposed Cerenkov fiber-optic dosimeter can be developed as a useful dosimeter to accurately obtain dose information prior to conducting radiotherapy. © 2013 IEEE.
- 제목
- Feasibility study on development of Cerenkov fiber-optic dosimeter for radiotherapy application
- 저자
- Yoo, W.J.; Shin, S.H.; Han, K.-T.; Jeon, D.; Hong, S.; Kim, S.G.; Sim, H.I.; Jang, K.W.; Cho, S.; Park, B.G.; Lee, B.
- 발행일
- 2013
- 학회명
- 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2013
- 개최지
- Osaka
- 학회 개최일
- 2013-07-03 ~ 2013-07-07