잔류염소 측정용 전기화학센서의 유효성

Effectiveness of the Electrochemical Sensor for the Free Chlorine Measurement

초록

Sodium hypochlorite is used worldwide as a water disinfectant and in bleaching agent. Sodium hypochlorite applied to water initially undergoes hydrolysis to form free chlorine consisting of hypochlorous acid(HOCl) and hypochlorite ion($OCl-). For free chlorine determination, an electrochemical method is simple due to the electroactivity of free chlorine; it measures current and is free of most reagents. Amperometric free chlorine sensor has been developed with gold (Au)-based electrode. The 3-electrode free chlorine sensor whose working and counter electrodes were Pt exhibited excellent response to HClO at +400mV vs. Ag/AgCl/sat. KCl. In addition, the use of a pH error correction algorithm provided a reliable measurement of residual free chlorine in water sample without any pretreatment in the normal pH range(pH 6~8) of municipal water supply. The free chlorine sensor installed in on-line monitoring system could be used to continually monitor the level of residual free chlorine in real samples.

키워드

Free residual chlorineElectroanalytical methodTree electrode systemAmperometryGold disk electrode유리잔류염소전기분석화학 방법3 전극계전류법금 전극
제목
잔류염소 측정용 전기화학센서의 유효성
제목 (타언어)
Effectiveness of the Electrochemical Sensor for the Free Chlorine Measurement
저자
김홍원정남용
DOI
10.7735/ksmte.2012.21.5.720
발행일
2012-10
저널명
한국생산제조학회지
21
5
페이지
720 ~ 725