상세 보기
감마선 분포 측정을 위한 섬광필름 기반의 감마 영상 검출기 제작 및 성능평가
- 신상훈;
- 유욱재;
- 장경원;
- 조승현;
- 이봉수
초록
As a feasibility study on development of a gamma imaging probe, we developed a scintillating film-based gamma imaging detector that can obtain scintillation images with information of gamma-ray distribution. The scintillating film-based gamma imaging detector was composed of a sensing probe, an image intensifier, and a beam profiler. To detect and transmit scintillation image, the sensing probe was fabricated by coupling a scintillating film, a fiber-optic image conduit, and a fiber-optic taper, consecutively. First, the optical images of USAF 1951 resolution target were obtained and then, modulation transfer function values were calculated to test the image quality of the sensing probe. Second, we measured the scintillation images according to the activity of the 137Cs and the distance between the surface of 137Cs and the distal-end of sensing probe. Finally, the intensities of scintillating light as functions of the activity and the distance were evaluated from the region of interest in the scintillation image. From the results of this study, it is expected that a fiber-optic gamma imaging detector can be developed to detect gamma-rays emitted from radiopharmaceuticals during radioimmunoguided surgery.
키워드
- 제목
- 감마선 분포 측정을 위한 섬광필름 기반의 감마 영상 검출기 제작 및 성능평가
- 제목 (타언어)
- Fabrication and Performance Evaluation of a Scintillating Film-based Gamma Imaging Detector to Measure Gamma-ray Distribution
- 저자
- 신상훈; 유욱재; 장경원; 조승현; 이봉수
- 발행일
- 2015-05
- 저널명
- 센서학회지
- 권
- 24
- 호
- 3
- 페이지
- 202 ~ 207